Medicine and medical device safety: 25 November 2026 review

This review covers 16 verified official safety communications and actions identified in the 24–25 November monitoring window. Document dates, regulatory decision dates and website publication dates are distinguished. A new posting does not necessarily represent a newly emerging signal.

FDA

1. Par Health: recall of 46 lots of six injectable medicines

FDA posted Par Health’s 23 November announcement on 24 November: 46 lots of Brevital Sodium, Coly-Mycin M Parenteral, Dantrium Intravenous, doxycycline, ethacrynate sodium and micafungin are recalled because of particulate matter. The company describes potentially life-threatening vascular consequences and reports no related adverse events to date. Wholesalers and hospital pharmacies should stop using and distributing affected lots and arrange returns. Check the official table for the exact lots and expiry dates.

Official posting: 2026-11-24. Document/action: 2026-11-23. Reference: not stated.

Official source

FDA

2. GSMS: one lot of ibuprofen 800 mg recalled

FDA posted the GSMS recall on 24 November. It covers only ibuprofen 800 mg lot GS069303, NDC 51407-841-05, expiry 01/2030, because of potential aluminium particles on tablet surfaces. The announcement describes minor gastrointestinal irritation or abrasion and a remote likelihood of a related adverse event; none has been reported. Consumers are instructed to stop using the affected lot and immediately contact a physician or pharmacist.

Official posting: 2026-11-24. Document/action: 2026-11-23. Reference: not stated.

Official source

FDA

3. Green Lumber: lot GL01 recalled for undeclared tadalafil

FDA published the 19 November announcement on 24 November. Green Lumber Natural Fuel For Men lot GL01, expiry 09/27, is recalled for undeclared tadalafil, which can dangerously lower blood pressure when combined with nitrates. Earlier lot exclusions or packaging comparisons should not be treated as safety assurances. Stop using the affected lot and contact the company about return arrangements. The notice identifies other current lots as outside this recall.

Official posting: 2026-11-24. Document/action: 2026-11-19. Reference: not stated.

Official source

MHRA

4. Toradol: missing warning in patient leaflets

MHRA issued a Class 4 caution for specified Toradol 30 mg/mL ketorolac batches because patient leaflets omit gastrointestinal anastomotic leakage safety information. Eight distributed batches comprise 91,134 packs; another batch was not distributed. This is not a recall: solution quality and efficacy are unaffected and it remains suitable when used with current product information. Before administration, consider recent or planned gastrointestinal surgery, verify this history with the patient where appropriate, and consult the current SmPC.

Official posting: 2026-11-24. Document/action: 2026-11-24. Reference: EL(26)A/50.

Official source
· Official document

MHRA

5. Brintellix: 5 mg tablets in cartons labelled 10 mg

MHRA’s Class 2 recall concerns one Quadrant Pharmaceuticals parallel-import batch of Brintellix: 2856920/4245Z, secondary batch MAX 4245Z, expiry 08/2029. Some cartons labelled 10 mg contain 5 mg tablets. Underdosing may reduce efficacy and worsen depression. Pharmacies should identify and quarantine remaining affected stock and contact patients who may have received it, following the full notice. This finding does not apply to every Brintellix batch. The notice requires action within 48 hours.

Official posting: 2026-11-24. Document/action: 2026-11-24. Reference: EL(26)A/49.

Official source
· Official document

Health Canada

6. QinFlow Warrior Lite: incorrectly assembled internal shield

Health Canada published RA-83001 on 24 November for specified Warrior Lite Base Units QIF-03; the recall date is 11 November. Investigation of field complaints involving disposable-unit thermal deformation found an incorrectly assembled mechanical shield on the circuit board. Harsh conditions or significant impact may detach the shield and cause an electrical short. This is a Type II action. Obtain the exact affected serial numbers and instructions from the manufacturer.

Official posting: 2026-11-24. Document/action: 2026-11-11. Reference: RA-83001;QIF-03.

Official source

Health Canada

7. CARESITE Micro: correction for possible valve closure failure

Health Canada posted this Type II notice on 24 November; the correction date is 9 November. B. Braun is correcting specified CARESITE Micro Luer Access Devices and extension-set lots because the piston may remain depressed after male-luer disconnection, preventing valve closure and causing leakage. Check the complete model/lot table and obtain manufacturer instructions. The notice describes a correction, rather than a blanket withdrawal of all these devices.

Official posting: 2026-11-24. Document/action: 2026-11-09. Reference: RA-82997.

Official source

Health Canada

8. Tavneos: Canadian patient transition and market withdrawal

Health Canada found insufficient evidence to support continued Tavneos authorization for its approved indication and announced Canadian market withdrawal. Controlled distribution for existing patients will continue until 22 May 2027 to support transition to alternatives; treatment should not be initiated in new patients. Continue liver-function monitoring during the transition and follow the updated permanent-discontinuation criteria, including suspected vanishing bile duct syndrome. Patients should not stop treatment without consulting their healthcare professional. The official communication contains the full laboratory thresholds.

Official posting: 2026-11-24. Document/action: 2026-11-24. Reference: RA-82995.

Official source

EMA

9. Landiolol: severe hypoglycaemia with sulfonylureas

EMA’s 24 November PSUSA publication identifies a reasonably possible causal link between concomitant landiolol/sulfonylurea use and severe hypoglycaemia. Sections 4.4/4.5 and the leaflet will be amended, including careful blood-glucose monitoring in diabetic patients. The CMDh position was adopted in October 2026; translations are due 29 November 2026 and variation submissions 28 January 2027. The benefit-risk balance is unchanged subject to the proposed information changes.

Official posting: 2026-11-24. Document/action: October 2026 (CMDh position). Reference: PSUSA/00010570/202602.

Official source
· Official document

EMA

10. Etoposide: interaction with mitotane and reduced efficacy

EMA published the etoposide PSUSA outcome on 24 November. Mitotane is to be added to the interaction information: CYP3A4 induction can increase etoposide clearance and reduce exposure and efficacy. A clinically relevant interaction is considered reasonably possible. SmPC section 4.5 and the leaflet will change. The October CMDh position sets 29 November 2026 for translations and 28 January 2027 for variation submissions; benefit-risk remains unchanged subject to these amendments.

Official posting: 2026-11-24. Document/action: October 2026 (CMDh position). Reference: PSUSA/00001333/202602.

Official source
· Official document

EMA

11. Strimvelis: warnings and administration information updated

EMA’s procedural record, published 24 November, documents the 15 October opinion and 20 November Commission decision. It records a hepatoblastoma case and mobilisation information added to SmPC section 4.4; section 4.2 changes for mobilised peripheral blood, including removal of the upper 20 × 10⁶ CD34+ cells/kg dose limit; and consideration of interactions involving mobilisation and myeloablative-conditioning medicines. These are the changes explicitly documented in the procedural record, which does not by itself establish that treatment caused the tumour.

Official posting: 2026-11-24. Document/action: 2026-11-20. Reference: EMA/VR/0000347859.

Official source
· Official document

ANSM

12. OptiCross IVUS: flushing and air-management instructions

ANSM posted Boston Scientific’s 22 October letter on 24 November. It clarifies catheter preparation and air management: do not flush an OptiCross IVUS catheter inside the patient when trapped air is suspected; remove it before additional flushing. The notice follows air-embolism reports. The manufacturer’s investigation did not link these events to catheter design or manufacturing. Devices may continue to be used with the updated instructions; there is no recall or special management requirement for previously treated patients.

Official posting: 2026-11-24. Document/action: 2026-10-22. Reference: R2633865;97661064-FA.

Official source
· Official document

ANSM

13. Shockwave Reducer: stop-use and return action

ANSM posted Shockwave Medical’s 2 November notice on 24 November. Specified Reducer lots, references R1385CEV/R1385GBV, are subject to a stop-use and return action. The reason is a safety signal still under investigation; additional information is needed to determine its potential health impact. The notice does not establish a particular injury mechanism. Continue standard monitoring of patients with an implanted Reducer and consult the official lot table.

Official posting: 2026-11-24. Document/action: 2026-11-02. Reference: R2633204;QA-FSCA-2026-65.

Official source
· Official document

ANSM

14. Getinge: replacement of specified sevoflurane vaporizers

ANSM posted Getinge’s 5 October revision 1 notice on 24 November. Specified Flow-i/c/e sevoflurane vaporizers were manufactured outside specifications, with potential long-term risks involving degradation products and inhalation. No patient or operator injuries have been reported. Replacement is targeted by 5 October 2027. Continued use is permitted under the stated handling precautions, including the 30-day sevoflurane retention limit and instructions for external transport. Check the official references and serial-number ranges.

Official posting: 2026-11-24. Document/action: 2026-10-05. Reference: R2630935;MX-9641Rev1;TW1569362.

Official source
· Official document

ANSM

15. Meta Glasses: software update to prevent overheating

ANSM’s 24 November posting carries the 9 November notice for Adventurer, Fury, Starfire, Nova and Capri glasses running software below version 130. Sweat or moisture can cause corrosion, a short circuit and local heating with temporary skin injury; no serious incidents have been reported. Install version 130 or later before further use. Ray-Ban Meta and Oakley Meta are explicitly outside this notice.

Official posting: 2026-11-24. Document/action: 2026-11-09. Reference: R2634041;000018.

Official source
· Official document

TGA

16. Activate Immune Complex: recall linked to anaphylaxis risk

TGA lists the 23 November Class I consumer recall of all unexpired Activate Immune Complex batches; ARTG 377233 is cancelled. The action cites sustained anaphylaxis reports for Andrographis-containing medicines despite warnings. Consumers are instructed to immediately discontinue use and dispose of the product appropriately.

Action commenced: 2026-11-23. Reference: RC-2026-RN-00954-1;ARTG377233.

Official source

Professional applicability

For QPPV, LQPPV and LCPPV work, assess applicability to local responsibilities and the relevant portfolio. Follow-up may include checking affected batches/devices, current product information, approved risk-minimisation measures and required action records. Role names and duties differ between jurisdictions.