Medicine and medical device safety: 27 November 2026 review
- 27-11-2026
- Posted by: Eldaniz Hasanov
- Category: Regulatory Updates
Medicine and medical device safety: 27 November 2026 review
A review of newly identified official communications checked during 25–27 November 2026. Document and corrective-action commencement dates may be earlier and are stated separately. Communications from different jurisdictions about the same event do not represent multiple new safety signals.
Сінрайз / Cinryze; human C1 esterase inhibitor
Ukraine SEC · Card / observation: 2026-11-27; document / action commencement: 2026-11; EXA/UA/CIN/0007
Ukraine’s SEC published Cinryze educational materials for healthcare professionals, patients and carers, covering preparation, administration, training for home use and a patient diary. This is a risk-minimisation package, rather than an announcement of a newly identified safety signal.
PROSTAP PD DCS 1.88 mg; leuprorelin acetate
MHRA · Card / observation: 2026-11-25; document / action commencement: 2026-11-25; EL(26)A/52; DMRC-41456133
MHRA states that PROSTAP PD DCS 1.88 mg must be administered subcutaneously only. Some administration instructions and syringe labels incorrectly permit intramuscular use. Batches 523373 and already expired 518961 are in scope. This is a Class 4 notification; no confirmed incorrect-route administrations or associated clinical outcomes have been identified for the UK presentation. Treating teams should review administration records.
Semaglutide
Health Canada · Card / observation: 2026-11-25; document / action commencement: 2026-11-25; RA-83010; SSR1795630406073
Health Canada found a possible link between semaglutide and very rare non-arteritic anterior ischaemic optic neuropathy (NAION). Most Canadian product monographs already include the warning; remaining updates are ongoing. Sudden vision loss or rapidly worsening eyesight requires urgent ophthalmological assessment; semaglutide should be discontinued if NAION is confirmed. Other GLP-1 receptor agonists remain under assessment.
Magnesium-containing parenteral medicines
BfArM · Card / observation: 2026-11-26; document / action commencement: 2026-11-26
BfArM is harmonising strength descriptions for magnesium-containing parenteral medicines to mmol/mL of Mg²⁺ to reduce medication errors. Old and new labels may coexist during the transition. Healthcare organisations should review prescribing records, protocols, labels and electronic systems. The communication describes a previously reported serious administration error.
Valproate de Sodium Arrow 200 mg/mL
ANSM · Card / observation: 2026-11-25; document / action commencement: not stated; MED26/B023
ANSM reports a healthcare-facility recall of Valproate de Sodium Arrow 200 mg/mL batches 1047, 1048, 1056 and 1057. Two sealed packs from batch 1057 contained an oral syringe labelled “Bétaméthasone 0.05%” and graduated in drops rather than milligrams. This could cause valproate overdose, particularly in young children. No related pharmacovigilance cases had been reported at the time of the notice.
Humexlib Etat Grippal 500/200/25 mg
ANSM · Card / observation: 2026-11-26; document / action commencement: not stated; No alert number assigned
ANSM reports a precautionary recall of Humexlib Etat Grippal 500/200/25 mg batches 1316 and 1317 directly from the pharmacies supplied. Braille incorrectly states “sugar-free”, although the medicine contains sugar. The laboratory reported no related pharmacovigilance cases at the time of the notice.
Medicines; AI-generated advice
ANSM · Card / observation: 2026-11-25; document / action commencement: not stated
ANSM warns about medication errors after patients followed AI-generated advice. Eight reports were recorded in France during 2024–2026, including one hospitalisation; this count is not an incidence estimate. The regulator advises checking official sources and discussing treatment changes with a healthcare professional.
Nitrosamine impurities
BfArM · Card / observation: 2026-11-27; document / action commencement: 2026-11-03; EMA/409815/2020 Rev.24
BfArM published a notice about updated EMA/CMDh guidance on nitrosamine impurities. The verified attachment is EMA/409815/2020 Revision 24, dated 3 November 2026. The national notice is dated 27 November; this is distinct from the document date and does not establish a new product-specific signal.
PEDMARQSI; sodium thiosulfate 80 mg/mL
MHRA · Card / observation: 2026-11-26; document / action commencement: 2026-11-17
The November MHRA roundup links a PEDMARQSI 80 mg/mL letter requiring visual inspection of each vial on receipt and before administration, and mandatory pre-administration or in-line filtration in accordance with the letter. These measures address possible particles; the primary letter provides the technical instructions. Sterility is unaffected and the letter identifies no infection risk. Clinical effectiveness of filtration for this issue has not been validated; individual benefit-risk assessment is required. The letter states that no new safety signal has been identified.
Warrior Lite Base Unit
ANSM · Card / observation: 2026-11-26; document / action commencement: 2026-11-11; FSFR 2026-01 / FSHU 2026-01
ANSM published a Warrior Lite Base Unit notice: a poorly secured internal shield may cause a short circuit and local overheating, creating burn risks. Equipment inspection, quarantine and the manufacturer’s specified measures apply.
HAMILTON-C6
ANSM · Card / observation: 2026-11-26; document / action commencement: 2026-11-17; FSCA-2026-11-06
ANSM published a notice for specified HAMILTON-C6 ventilators: a turbine issue may cause loss of mechanical ventilation. Immediate backup ventilation is required if the relevant failure occurs; the letter lists affected devices and corrective actions.
3Dimensions / Selenia Dimensions
ANSM · Card / observation: 2026-11-26; document / action commencement: 2026-11-10; MISC-12191-FRA-901 Rev.001; FA-00314
ANSM published Hologic’s notice on the vertical-motion mechanism of 3Dimensions and Selenia Dimensions systems. Wear can cause unexpected movement or C-arm descent. Six-monthly inspections are required; use should stop and service support should be contacted if jerking, jamming or unusual noise occurs.
LUNA X5 / LUNA IPX4
ANSM · Card / observation: 2026-11-26; document / action commencement: 2026-11-06; FSN_DK01069005
ANSM published Winncare’s updated notice on locking the portable LUNA lifting unit. Incorrect locking may cause a patient or equipment fall. Users experiencing any locking difficulty should immediately stop use and contact the manufacturer.
BeneVision WorkStation / ViewStation
ANSM · Card / observation: 2026-11-25; document / action commencement: 2026-11-18; CP2610-JH06304
ANSM published Mindray’s notice on possible BeneVision workstation disconnection from CentralStation. Bedside monitoring and alarms continue, but workstation data may disappear and delay intervention. Bedside checks and a software update are specified.
Xtrans / DPT-6000 pressure measurement sets
ANSM · Card / observation: 2026-11-25; document / action commencement: 2026-11-10; FSN_26-09-098; FSCA_26-09-098
ANSM published CODAN’s notice on excessive flush flow in specified Xtrans/DPT-6000 pressure measurement sets caused by an assembly defect. An undetected defect may cause excessive administration of flush solution. Manufacturer checks and instructions apply.
FRED 2 anti-fog kit
ANSM · Card / observation: 2026-11-25; document / action commencement: 2026-11; FA00013
ANSM published Medtronic’s notice for specified FRED 2 lots: solution leakage onto the Tyvek lid may compromise the sterile barrier and create a biocontamination risk. The notice provides clinical mitigations and does not request product retrieval.
HeartSine Samaritan PAD 350P/360P/500P
Health Canada · Card / observation: 2026-11-25; document / action commencement: Recall 2026-11-23; RA-83012
Health Canada published a Type I recall for specified HeartSine Samaritan PAD devices. Oxidation of spring-loaded contacts may cause an intermittent or lost electrical connection to the Pad-Pak/Pediatric-Pak battery. Affected devices and required actions should be checked with the manufacturer.
Novum IQ Syringe Pump
Health Canada · Card / observation: 2026-11-25; document / action commencement: Recall 2026-11-19; RA-83003
Health Canada reports a correction for Novum IQ Syringe Pump models 40800BAXL, 40800BAX and 40800BAXRP. A software issue may occur during a Volume/Time multi-syringe infusion when an empty syringe is replaced with a different brand or size while a volume remains to be infused. Users should obtain further instructions from the manufacturer.
Lazcluze / lazertinib
Swissmedic · Card / observation: 2026-11-26; document / action commencement: 2026-11-02; RMP version 3.0
Swissmedic posted the Lazcluze RMP summary, version 3.0 dated 2 November. It lists venous thromboembolic events with amivantamab combination treatment and hepatotoxicity as important identified risks. This is a newly posted summary; no new clinical signal is inferred without a prior-version comparison.
Dysport; Darzalex; Imbruvica
ANVISA · Card / observation: 2026-11-25; document / action commencement: not stated; RE 4.623/2026
Anvisa reports seizure of counterfeit Dysport batch 039228 and irregular Darzalex/Imbruvica batches. The notice distinguishes counterfeits from foreign-market products that do not match the Brazilian registration. Imbruvica 140 mg batch 06A02V is identified as counterfeit; these findings must not be generalised to the rights holders’ portfolios.
SmartPort Implantable Ports
Health Canada · Card / observation: 2026-11-25; document / action commencement: Recall 2026-11-11; RA-83002
Health Canada reports a recall of specified SmartPort Implantable Ports: incorrectly sized components may prevent a secure port-to-catheter connection or catheter advancement through the introducer sheath. The card was published on 25 November; the recall commenced on 11 November. Affected lots and models are listed in the official card.
Medline beds, rails, mattresses and accessories
Health Canada · Card / observation: Updated 2026-11-25; original 2025-12-30; document / action commencement: Recall 2026-11-10; RA-81440
Health Canada’s Medline beds, rails, mattresses and accessories record, updated on 25 November, identifies electrical-system fire risks and potential entrapment involving non-Medline accessories. The original publication is dated 30 December 2025 and the recall date is 10 November 2026. The updated card is not presented as a new signal arising on 25 November.
Hamilton/47244/Ventilators
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-25; Medsafe 38060
Medsafe action 38060, commenced on 25 November, concerns premature HAMILTON-C6 blower wear that may cause loss of mechanical ventilation even after a successful pre-use test. Immediate alternative ventilation is required on failure. The New Zealand serial-number list differs from the French notice; users should consult the source for their market.
EYE-LIGHT
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-24; Medsafe 37961
Medsafe reports updated EYE-LIGHT/MY MASK-E instructions: certain anatomical or ophthalmic conditions may reduce the required skin-to-LED distance and cause excessive local heating. Action 37961 commenced on 24 November.
Corpuls 3 Defib including Accessories and Upgrades
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-24; Medsafe 38058
Medsafe reports Corpuls 3/3T concerns involving continuing reports of failure to deliver a shock and spontaneous reboots. Additional measurement and data-transmission issues remain under investigation. Action 38058 commenced on 24 November; the card identifies models, software and measures.
Stryker Triathlon Hinge Trays
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-23; Medsafe 38031
Medsafe reports revised reprocessing instructions for specified Stryker Triathlon Hinge Trays. Sterilisation in blue wrap requires 40 rather than 30 minutes of drying; the shorter time could leave residual moisture. Action 38031 commenced on 23 November; no product return is required.
Corpuls 3 Defib including Accessories and Upgrades
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-23; Medsafe 38039
A separate Medsafe Corpuls 3/3T action, 38039 commenced on 23 November, concerns Bluetooth interruption when printing a resting ECG. Display and controls may be unavailable for 5–25 seconds, potentially delaying lifesaving treatment in an emergency. Device modification is specified.
G3R vital signs monitor with mobile stand
Medsafe · Card / observation: 2026-11-27; document / action commencement: 2026-11-19; Medsafe 38022
Medsafe reports replacement of specified G3R monitor power cords: missing required insulation on active and neutral plug pins creates an electric-shock hazard during insertion or removal. Action 38022 commenced on 19 November.
These communications may support QPPV, LQPPV and LCPPV review of locally applicable measures and product information. Role names and responsibilities vary between jurisdictions.